Description: ESA-E Spectrum Analyzer Tracking Generator
Manufacturer | Keysight Technologies |
---|---|
Series | ESA Series |
Part Number | E4407B |
Need to purchase this item?
HP Agilent Keysight E4407B ESA-E Spectrum Analyzer
The E4407B Spectrum Analyzer has a frequency range between 9 kHz to 26.5 GHz. It also has a VGA TFT active matrix display, which makes it suitable for different environmental applications. The E4407B Spectrum Analyzer offers reduced emission and a Bluetooth FM demodulator. It also includes frequency reference for high stability and noise figure measurements. The model has a continuous phase lock synthesizer, which functions over the entire swap and aids in achieving high frequency accuracy. This helps with gaining stability and incorporates features to enable users for amplitude modification. It eliminates effects related to frequency from the signal.
The E4407B Spectrum Analyzer can also have optional filters. These digital resolution bandwidth filters process a frequency range of 10 Hz to 300 Hz. These digital RBW filters sweep quickly, around 220 times faster than their analog counterparts. This provides high improvement in terms of sensitivity.
The E4407B Spectrum Analyzer offers features that simplify various testing requirements. It offers many limit lines with margins that help to conduct testing in the production line. This feature enables users to go through all tests in detail with convenience. The Keysight E4407B is also integrated with a built-in frequency counter. This counter has a resolution of 1 Hz. It assists with high precision in the measurement of individual signals. Features like ACP, occupied BW, and emission BW are also available in this product. These features are fully controlled through one-button measurements and help the user achieve swift and repeatable outputs.
Model | E4407B ESA-E |
---|---|
Manufacturer | HP Agilent Keysight |
Series | ESA-E |
Type | Spectrum Analyzer |
Frequency Range | 9 kHz to 26.5 GHz |
Marker peak search | 300 ms |
Counter Resolution | Selectable from 1 Hz to 100 kHz |
Resolution Bandwidth Accuracy | 1 to 300 Hz ±10% |
Status | One Year |
E4407B ESA-E Manual | User Manual and Maintenance Manual Available for Download |
Application | Industrial |
HP Agilent Keysight E4407B ESA-E Spectrum Analyzer- FAQ:
Question:
Why didn’t my E4407B ESA-E arrive with every personality that I ordered installed already?
Answer:
If more than 2 or 3 personalities were ordered, the memory space may not have been large enough to accommodate.
Question:
Why is the internal preamp necessary to perform noise figure measurements on the E4407B ESA-E Spectrum Analyzer?
Answer:
It is required to order the Internal Preamplifier Option 1DS because all model specifications below 3 GHz need this option installed. It will help decrease the total NF measurement uncertainty.
Question:
Why is my E4407B ESA-E model making a chirping noise?
Answer:
There is an automated realignment feature during every retrace. While realignments are occurring, the analyzer switches signals in the IF using multiple relays. The relays cause a chirping or clicking noise, which is normal for the model.
Question:
Is there a viewing hood for the Keysight E4407B ESA-E?
Answer:
No, the display used in the E4407B has a huge viewing angle; the contrast can be changed if direct sunlight is an issue.
Question:
Can I perform phase noise measurements on my E4407B Spectrum Analyzer while using Option 226 when the spectrum analyzer is in external mixing mode (Option AYZ)?
Answer:
Automated phase noise measurements using Option 226 can be performed within the base frequency of the E4407B only. They cannot be performed while in external mixing mode.